Mineral Oil is added to Crystal Polystyrene to increase its flexibility. Precise control of the amount of mineral oil
used is necessary to ensure the correct performance properties are obtained reproducibly. NMR can provide a rapid and
effective means of monitoring the mineral oil content in crystal polystyrene.
Method
NMR has a number of advantages over other techniques:
- NMR is very stable over the long term and rarely needs calibration adjustment.
- The measurement time is short (typically two seconds).
- The NMR technique is non-destructive, so the same sample may be measured several times before being analysed by
other techniques.
- NMR is insensitive to air voids between polystyrene granules.
- Measurement precision is good compared to wet chemical methods.
- NMR does not require the use of hazardous solvents.
- Both weighing and non-weighing methods are available for this application.
Calibration and Results
Five samples of crystal polystyrene were selected for the analysis of mineral oil content. Each sample was weighed
into a tared 26mm glass tube and preconditioned for an hour at 105°C, before being placed in an MQC+ for analysis.
The graph in Figure 1 illustrates the calibration of the NMR measurements against mineral oil contents determined by
a reference method for these samples. The quality of the calibration exhibited in Figure 1 is excellent, as indicated
by the correlation coefficient of 1 and the standard deviation of 0.07.

Fig. 1 — Calibration of NMR data and reference values for oil content. Correlation Coefficient: r = 1; Standard
Deviation = 0.07.
Recommended Instrument
The MQC+ with a 0.55 Tesla (23 MHz hydrogen resonance) magnet, fitted with a 26mm diameter (14ml sample) probe is a
suitable instrument for this application. The Oil in Crystal Polystyrene package consists of:
- MQC+ with a built-in computer operating Microsoft® Windows® 10 (no separate PC is required).
- MultiQuant software including RI Calibration, RI Analysis, and
the EasyCal ‘Oil in Crystal Polystyrene’ application.
- 26mm glass tubes.
- User Manual.
- ‘Oil in Crystal Polystyrene’ method sheet.
In addition to this package you will also require:
- A dry heater and aluminium block with 26mm holes for sample conditioning at 105°C.
- A precision balance.
The instrument offers multiple advantages over others on the market:
- High signal sensitivity.
- Small benchtop footprint.
- Low maintenance.
- Recyclable sample tubes, lowering consumable costs.
- Minimal sample preparation.
Note: Other instruments/packages are available for the analysis of larger or smaller sample
quantities. Please contact Oxford Instruments for further details.