Fluorides are used in many industrial processes, of which the production of hydrochloric acid and smelting are two
examples. It is important to regulate these processes, so that the yield of product is maximised and the emission of
toxic fluoride gases prevented. The MQC+ is capable of accurately and rapidly measuring the fluoride
content of solids in powder form.
Method
Conventional methods of measurement include titrometry, ion selective electrode, and ion chromatography. All of these
methods are cumbersome, time consuming and require skilled chemists. Furthermore, sample preparation may often require
the use of hazardous chemicals which require disposal.
The MQC+ provides a rapid and effective alternative method for determining fluoride content, and
offers a number of advantages over secondary methods:
- It can be calibrated to cover a concentration range from 0 to 100%.
- It can be calibrated using only 6 samples.
- NMR is very stable over the long term, so calibrations will rarely require adjustment.
- NMR is insensitive to air voids between grains of powder.
- The measurement time is short (typically 32 seconds).
- The NMR technique is non-destructive, so samples analysed may be re-used.
Samples are simply loaded into pre-tared glass vials, weighed, conditioned, then inserted into the instrument which
detects the sample, automatically starting the NMR analysis. The instrument returns the fluoride content values in
less than one minute.
Calibration and Results
Eleven artificial standards of Calcium Fluoride ranging from 1 to 100% CaF2 content were made up for analysis of
fluoride content. Each sample was weighed into a tared 26mm glass tube and preconditioned for 20 minutes at 40°C,
before being placed in the MQC+ for analysis. The results of this analysis are seen in Figure 1. The
quality of the calibration generated is good, especially as such a large range of fluoride is analysed.

Fig. 1 — Calibration of NMR data for fluoride content of calcium fluoride samples. Correlation Coefficient: r =
1.00; SD = 0.41
Table 1 shows that the repeatability (or precision) of the NMR measurement of the same sample is excellent.
| Repeat number |
Fluoride Content (%) |
| 1 |
24.8 |
| 2 |
24.7 |
| 3 |
24.8 |
| 4 |
24.7 |
| 5 |
24.8 |
| 6 |
24.8 |
| 7 |
24.9 |
| 8 |
24.5 |
| 9 |
24.9 |
| 10 |
24.8 |
| SD: |
0.116 |
Table 1: Repeatability of NMR Measurement of Calcium Fluoride samples.
Recommended Instrument Configuration
The MQC+ with a 0.55 Tesla (22 MHz fluorine resonance) magnet, fitted with a 26mm diameter (14ml
sample) probe is a suitable instrument for this application. The Fluoride Content in Powder package comprises:
- MQC+ with a built-in computer operating Microsoft® Windows® 10 (no separate PC is required).
- MultiQuant software including RI Calibration, RI Analysis, and
the EasyCal ‘Fluoride in Powder’ application.
- 26mm diameter glass tubes.
- User Manuals
- ‘Fluoride in Powder’ method sheet.
In addition to this package you will also require:
- A dry heater and aluminium block with 26mm holes for sample conditioning at 40°C.
- A precision balance.
The instrument offers advantages over others on the market:
- High signal sensitivity.
- Small benchtop footprint.
- Low maintenance.
- The sample tubes are recyclable, lowering consumable costs.
- Minimal sample preparation.